Citation: | YANG Liu, GAO Hui-li, WANG Yin-fu, WANG Hai-jiao. DETERMINATION OF NINETEEN TRACE ELEMENTS IN HIGH PURITY SILVER BY ICP-AES[J]. Geology and Resources, 2019, 28(1): 95-97. |
Nitric acid is used to dissolve the high purity silver ingot, with ICP-AES to determine the 19 trace elements as Al, Ca, Cd, Co, Cr, Cu, Fe, Mg, Mn, Ni, Pb, Au, Pd, Pt, Rh, Si, Sn, Te and Zn within. To obtain accurate experimental results, select appropriate instrument conditions and analysis spectral lines, then measure the curves to match the silver content. The recovery and accuracy of results prove to be satisfactory. Silver is separated as AgCl precipitate and the trace element contents in the supernatant is determined. The results show that the silver precipitation has little effect on the determination. Therefore, the test can be achieved without the matrix matching of silver. This method can thus eliminate silver interference and reduce sample testing cost.
[1] | 韩永志, 编.标准物质手册[M].北京:中国计量出版社, 1998:1-12. |
[2] | 侯列奇, 李洁, 王树安, 等.电感耦合等离子体原子发射光谱法测定高纯银中25种微量元素[J].分析实验室, 2007, 26(增):363-366. |
[3] | 何春羚.电感耦合等离子体原子发射光谱法测定铝合金中的硼[J].中国金属通报, 2017, 6:60-61. |
[4] | 宋丽华, 郝原芳, 高慧莉.地质样品中金的测定方法研究[J].地质与资源, 2014, 23(6):577-579. doi: 10.3969/j.issn.1671-1947.2014.06.016 |