Citation: | XIONG Ying, WU Zheng, DONG Ya-ni, PEI Ruo-hui, LIU Xiao-Yan, WU He, ZHANG Yan. Determination of Rhenium in Copper-Lead-Zinc Ore by Inductively Coupled Plasma-Mass Spectrometry with Closed Decomposition and Cation Exchange Separation[J]. Rock and Mineral Analysis, 2015, 34(6): 623-628. doi: 10.15898/j.cnki.11-2131/td.2015.06.004 |
Anion exchange resin has been applied to Re separation after acid digestion of Cu-Pb-Zn ores and Re was determined by Inductively Coupled Plasma-Mass Spectrometry(ICP-MS). ReO4- in solution was adsorbed by exchange resin and was then eluted from the resin by eluant. This method has a long process, which affects its stability and accuracy. Matrix elements are selected as the goal and 8 g of 732 cation exchange resin in 2% nitric acid medium for 1.5 h static adsorption is used. This method can effectively eliminate the influence of matrix elements on the ICP-MS determination of rhenium. Quartz was used as the blank and mixed with low Re samples. The acquired detection limit was 0.004μg/g, the precision(RSD) is less than 6%, and the recovery ranged from 96% to 104%. The 732 cation exchange resin under the acidic condition has the recovery of >95% for most matrix elements except tungsten and molybdenum, which reduced ionic strength and the concentration of interference elements. Rhenium can be directly determined by ICP-MS, which is simpler and faster than the anion exchange resin separation method.
[1] |
|
[2] |
|
[3] | |
[4] | |
[5] | |
[6] | |
[7] | |
[8] | |
[9] | |
[10] | |
[11] | |
[12] | |
[13] | |
[14] | |
[15] | |
[16] | |
[17] | |
[18] |
Effect of acidity, exchange adsorption time and amount of resin on separation effect